Electrical Safety Test
Electrical Safety Testing (EST) for Medical Equipment
Electrical Safety Testing (EST) is crucial to ensuring the safety and performance of medical equipment, especially in healthcare environments. Compliance with international and national electrical safety standards such as MS IEC 60364-7-710 and IEC 60601-1 is mandatory for all healthcare facilities. This includes active medical devices that must conform to MS IEC60601-1, IEC 60601-1 Collateral Standard, and IEC 60601-2.
Before any medical device is placed into clinical service, whether new or used, it must undergo thorough testing. This includes visual inspection, electrical safety tests, and performance tests. Any maintenance work involving the replacement of defective electrical components or disassembly of active medical devices must also be followed by an electrical safety test before the equipment is released for use.
Get in Touch
For more information about our Electrical Safety Testing (EST) services and to schedule a test appointment, contact us today.
IEC 60601-1 Compliance Test
The IEC 60601-1 standard defines the essential safety and performance requirements for medical electrical equipment to ensure the safety of patients and healthcare providers. It is essential that all medical equipment complies with this standard to prevent any electrical hazards during use.
Key Compliance Areas:
- Electrical Insulation: Verifies the integrity of insulation materials to avoid electrical leakage.
- Grounding: Ensures proper grounding to prevent electric shock.
- Leakage Current: Measures current leakage to ensure it stays below safe limits.
- Dielectric Strength: Assesses the electrical breakdown of insulation under high-voltage conditions.
Adherence to IEC 60601-1 ensures that medical devices meet high safety standards, protecting both patients and operators.
Before any medical device is placed into clinical service, whether new or used, it must undergo thorough testing. This includes visual inspection, electrical safety tests, and performance tests. Any maintenance work involving the replacement of defective electrical components or disassembly of active medical devices must also be followed by an electrical safety test before the equipment is released for use.
IEC 62353 Compliance Test
The IEC 62353 standard specifically addresses the electrical safety testing of medical electrical equipment used after the initial commissioning and throughout its operational life. It ensures that devices continue to operate safely throughout their life cycle, including regular maintenance checks.
Key Compliance Areas:
- Electrical Safety Testing (EST): Conducting tests for leakage current, grounding, and insulation resistance.
- Routine Testing: Continuous evaluation to identify potential safety risks.
- Accurate Results Interpretation: Understanding test results to take timely corrective actions.
Compliance with IEC 62353 is vital to ensure the ongoing safety of medical equipment during its lifespan and is a critical aspect of medical device maintenance.
IEC 61010 Compliance Test
The IEC 61010 standard outlines safety requirements for electrical equipment used in measurement, control, and laboratory environments. For medical devices used in diagnostic testing or laboratory settings, adherence to this standard is essential to guarantee safety during operation.
Key Compliance Areas:
- Safety During Operation: Ensures that electrical equipment for measurements and control does not pose any risks during usage.
- Overvoltage Protection: Safeguards the device and operators from overvoltage scenarios.
- Earthing and Insulation: Ensures proper earthing and insulation to prevent shock hazards.
Adhering to IEC 61010 ensures that diagnostic equipment and laboratory devices meet the required safety standards for safe and reliable operation.
Key Compliance Areas:
- Electrical Safety Testing (EST): Conducting tests for leakage current, grounding, and insulation resistance.
- Routine Testing: Continuous evaluation to identify potential safety risks.
- Accurate Results Interpretation: Understanding test results to take timely corrective actions.
Compliance with IEC 62353 is vital to ensure the ongoing safety of medical equipment during its lifespan and is a critical aspect of medical device maintenance.
3-Phase Medical Equipment
3-phase medical equipment is used in healthcare facilities to deliver high power for sophisticated machinery like CT scanners, MRI machines, and surgical equipment. These devices require comprehensive electrical safety testing to ensure reliable and safe performance.
Testing Process:
- Earth (Ground) Continuity Test: Ensures proper grounding to reduce shock risks.
- Insulation Resistance Test: Verifies that the insulation is intact, preventing electrical leakage.
- Dielectric Strength (Hipot) Test: Ensures insulation can withstand high voltages without breaking down.
- Leakage Current Test: Measures any electrical current leakage to ensure safety.
Safety Standards:
- IEC 60601-1: Addresses electrical safety for medical electrical equipment, focusing on patient and operator safety.
- IEC 61010: Sets safety requirements for electrical equipment used in measurement, control, and laboratory applications.
Why Compliance is Important
Compliance with IEC 60601-1 and IEC 61010 for 3-phase equipment guarantees the safe operation of critical medical devices. It helps prevent electrical hazards, ensuring a safe environment for both patients and healthcare professionals.